FDA Device recall Z-0512-2013

Siemens Healthcare ADVIA Centaur Theophylline 2 (THE02) 250 Test 08838753 50 Test 03009619 Intended Use For in vitro diagnostic use in the…

On 2012-12-19 the FDA classified a Class II recall of Siemens Healthcare ADVIA Centaur Theophylline 2 (THE02) 250 Test 08838753 50 Test 03009619 Intended Use For in vitro diagnostic use in the… by Siemens Healthcare Diagnostics, citing operating range of the ADVIA Centaur systems is 18¿C to 30¿C (64.4¿F to 86.0¿F), a change in room temperature may increase or decrease assay results, depending on the specific ass….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0512-2013
ClassificationClass II
Statusterminated
Initiated2012-11-14
Published2012-12-19
Last updated2026-09-13 11:53 UTC
CompanySiemens Healthcare Diagnostics
DistributionUS

Product

Siemens Healthcare ADVIA Centaur Theophylline 2 (THE02) 250 Test 08838753 50 Test 03009619 Intended Use For in vitro diagnostic use in the quantitative determination of theophylline in serum or plasma using the ADVIA Centaur and ADVIA Centaur XP systems.

Reason

Operating range of the ADVIA Centaur systems is 18¿C to 30¿C (64.4¿F to 86.0¿F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used.

Identifiers

code_infoAll lot codes