FDA Device recall Z-0492-2026
3gAllergy Specific IgE Universal Kit
- FDA Device
- Class II
- ongoing
- Published 2025-11-19
On 2025-11-19 the FDA classified a Class II recall of 3gAllergy Specific IgE Universal Kit by Siemens Healthcare Diagnostics, citing the barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument i….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0492-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-10-06 |
| Published | 2025-11-19 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Siemens Healthcare Diagnostics |
| Distribution | US |
Product
3gAllergy Specific IgE Universal Kit; Kit Component: 3gAllergy Specific IgE Adjustor Antibody (L2UNS1), 3gAllergy Specific IgE Control Antibody (L2UNS2); Siemens Material Number (SMN): 10380875; Catalog Number: L2KUN6;
Reason
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
Identifiers
| code_info | Siemens Material Number (SMN): 10380875; Model/Catalog: L2KUN6; UDI-DI: 00630414962269; Lot number: 169, 170, 171, 172; |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0492-2026%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.