FDA Device recall Z-0484-2022
GLOBAL UNITE STD STEM SZ 8 Part Number: 110008100
- FDA Device
- Class II
- terminated
- Published 2022-01-19
On 2022-01-19 the FDA classified a Class II recall of GLOBAL UNITE STD STEM SZ 8 Part Number: 110008100 by Depuy Orthopaedics, citing stems may have an undersized spigot bore diameter therefore separating the proximal body from the stem may not be possible without surgical delay and/or bone damage.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0484-2022 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2021-12-07 |
| Published | 2022-01-19 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Depuy Orthopaedics |
| Distribution | US |
Product
GLOBAL UNITE STD STEM SZ 8 Part Number: 110008100
Reason
Stems may have an undersized spigot bore diameter therefore separating the proximal body from the stem may not be possible without surgical delay and/or bone damage
Identifiers
| code_info | Lot # s 9793035, 9802134, 9802135, 9802136, 9802137, 9802138, 9802139, 9802140, 9802141, 9802142, 9802143, 9802144, 980…Lot # s 9793035, 9802134, 9802135, 9802136, 9802137, 9802138, 9802139, 9802140, 9802141, 9802142, 9802143, 9802144, 9802145, 9802146, 9802147, 9802148, 9802149, 9802150, 9802151, 9814515, 9817165, 9817166, 9817167, 9817168, 9817633, 9818331, 9824332, 9824334, 9824335, 9824336, 9824337, 9824338, 9824339, 9824340, 9824341, 9824342, 9824343, 9824344, 9824345, 9824346, 9824347, 9832804, 9851705, 9851706, 9851707, 9851708, 9862057, 9864147, 9874672, 9874674, 9874675, 9874676, 9874678, 9874682, 9874683, 9874686, 9874688, 9887446 GTIN: 10603295004042 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0484-2022%22