FDA Device recall Z-1267-2023
Attune Posterior (PS) Fixed Bearing (FB) Tibial Insert SZ5 7MM, Part Number 151640507
- FDA Device
- Class II
- ongoing
- Published 2023-03-29
On 2023-03-29 the FDA classified a Class II recall of Attune Posterior (PS) Fixed Bearing (FB) Tibial Insert SZ5 7MM, Part Number 151640507 by Depuy Orthopaedics, citing the subject product lots are being recalled because they received a higher than specified irradiation dose. This exceeds the validated range for exposure to gamma radiation of the….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1267-2023 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2023-02-15 |
| Published | 2023-03-29 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Depuy Orthopaedics |
| Distribution | US |
Product
Attune Posterior (PS) Fixed Bearing (FB) Tibial Insert SZ5 7MM, Part Number 151640507
Reason
The subject product lots are being recalled because they received a higher than specified irradiation dose. This exceeds the validated range for exposure to gamma radiation of these devices and may result in changes to the implant material properties.
Identifiers
| code_info | GTIN: 10603295049944; Lot: JN6613 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1267-2023%22