FDA Device recall Z-0483-2022
GLOBAL UNITE STD STEM SZ 6 Part Number: 11000610
- FDA Device
- Class II
- terminated
- Published 2022-01-19
On 2022-01-19 the FDA classified a Class II recall of GLOBAL UNITE STD STEM SZ 6 Part Number: 11000610 by Depuy Orthopaedics, citing stems may have an undersized spigot bore diameter therefore separating the proximal body from the stem may not be possible without surgical delay and/or bone damage.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0483-2022 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2021-12-07 |
| Published | 2022-01-19 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Depuy Orthopaedics |
| Distribution | US |
Product
GLOBAL UNITE STD STEM SZ 6 Part Number: 11000610
Reason
Stems may have an undersized spigot bore diameter therefore separating the proximal body from the stem may not be possible without surgical delay and/or bone damage
Identifiers
| code_info | Lot #: 9792595, 9793039, 9802121, 9813506. 9818509, 9830922, 9840359,9858603 GTIN: 10603295004011 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0483-2022%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.