FDA Device recall Z-0468-2016
Factor VIII Chromogenic Assay, REF/Catalog No. B4238-40, Siemens Material Number (SMN) 10445729, IVD. --- CLASSIFICATION NAME: Test, Qualit…
- FDA Device
- Class II
- terminated
- Published 2015-12-30
On 2015-12-30 the FDA classified a Class II recall of Factor VIII Chromogenic Assay, REF/Catalog No. B4238-40, Siemens Material Number (SMN) 10445729, IVD. --- CLASSIFICATION NAME: Test, Qualit… by Siemens Healthcare Diagnostics, citing factor VIII Chromogenic Assay Lot 44637 is showing an increased occurrence of irregular curves when used for measurement on Sysmex CA-1500, CA-7000, CS-2000i, CS-2100i, CS-5100 Sy….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0468-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-11-19 |
| Published | 2015-12-30 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Siemens Healthcare Diagnostics |
| Distribution | MA |
Product
Factor VIII Chromogenic Assay, REF/Catalog No. B4238-40, Siemens Material Number (SMN) 10445729, IVD. --- CLASSIFICATION NAME: Test, Qualitative and Quantitative Factor Deficiency
Reason
Factor VIII Chromogenic Assay Lot 44637 is showing an increased occurrence of irregular curves when used for measurement on Sysmex CA-1500, CA-7000, CS-2000i, CS-2100i, CS-5100 Systems. This could lead, under certain circumstances, to an erroneously increased result with a Factor VIII Chromogenic Assay.
Identifiers
| code_info | Lot Number 44637, exp. 2016-11-30 |
|---|