FDA Device recall Z-0468-2016

Factor VIII Chromogenic Assay, REF/Catalog No. B4238-40, Siemens Material Number (SMN) 10445729, IVD. --- CLASSIFICATION NAME: Test, Qualit…

On 2015-12-30 the FDA classified a Class II recall of Factor VIII Chromogenic Assay, REF/Catalog No. B4238-40, Siemens Material Number (SMN) 10445729, IVD. --- CLASSIFICATION NAME: Test, Qualit… by Siemens Healthcare Diagnostics, citing factor VIII Chromogenic Assay Lot 44637 is showing an increased occurrence of irregular curves when used for measurement on Sysmex CA-1500, CA-7000, CS-2000i, CS-2100i, CS-5100 Sy….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0468-2016
ClassificationClass II
Statusterminated
Initiated2015-11-19
Published2015-12-30
Last updated2026-09-13 11:53 UTC
CompanySiemens Healthcare Diagnostics
DistributionMA

Product

Factor VIII Chromogenic Assay, REF/Catalog No. B4238-40, Siemens Material Number (SMN) 10445729, IVD. --- CLASSIFICATION NAME: Test, Qualitative and Quantitative Factor Deficiency

Reason

Factor VIII Chromogenic Assay Lot 44637 is showing an increased occurrence of irregular curves when used for measurement on Sysmex CA-1500, CA-7000, CS-2000i, CS-2100i, CS-5100 Systems. This could lead, under certain circumstances, to an erroneously increased result with a Factor VIII Chromogenic Assay.

Identifiers

code_infoLot Number 44637, exp. 2016-11-30

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.