RecallRadar

FDA Device recall Z-0399-2022

SYNCHRON Systems Cholesterol (CHOL) Reagent, REF: 467825

On 2021-12-29 the FDA classified a Class II recall of SYNCHRON Systems Cholesterol (CHOL) Reagent, REF: 467825 by Beckman Coulter, citing cholesterol Reagent may fail with Out of Calibration Range Low (OCR Low) and/or Back to Back (B/B) errors. Low quality control recovery failures may be observed after passing cali….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0399-2022
ClassificationClass II
Statusongoing
Initiated2021-11-10
Published2021-12-29
Last updated2026-09-13 11:56 UTC
CompanyBeckman Coulter
DistributionUS

Product

SYNCHRON Systems Cholesterol (CHOL) Reagent, REF: 467825

Reason

Cholesterol Reagent may fail with Out of Calibration Range Low (OCR Low) and/or Back to Back (B/B) errors. Low quality control recovery failures may be observed after passing calibration. Brown particulates in compartment C of the reagent cartridge may be observed. Continued use may lead to, 1) failed QC and/or calibrations, 2) Erroneous low Cholesterol patient results.

Identifiers

code_infoLot: M106450, UDI: 15099590575113

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0399-2022%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.