FDA Device recall Z-0327-2026
VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7
- FDA Device
- Class II
- ongoing
- Published 2025-11-05
On 2025-11-05 the FDA classified a Class II recall of VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7 by Canon Medical System Usa, citing affected system contains components that do not meet the required distance between electrodes for compliance with the IEC 60601-1 standard for medical electrical equipment.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-0327-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-10-16 |
| Published | 2025-11-05 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Canon Medical System Usa |
| Distribution | NV, OH |
Product
VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7.
Reason
Affected system contains components that do not meet the required distance between electrodes for compliance with the IEC 60601-1 standard for medical electrical equipment.
Identifiers
| code_info | CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7. Serial Numbers: B7A2462002, B7A2522014. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0327-2026%22