FDA Device recall Z-1179-2026
Alphenix INFX-8000V, interventional fluoroscopic x-ray system
- FDA Device
- Class II
- ongoing
- Published 2026-02-04
On 2026-02-04 the FDA classified a Class II recall of Alphenix INFX-8000V, interventional fluoroscopic x-ray system by Canon Medical System Usa, citing it has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional syste….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-1179-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-12-19 |
| Published | 2026-02-04 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Canon Medical System Usa |
| Distribution | US |
Product
Alphenix INFX-8000V, interventional fluoroscopic x-ray system
Reason
It has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result, ceiling lateral movement may no longer be possible, abnormal noise may be generated from the suspension system and a sensor error may be displayed
Identifiers
| code_info | serial numbers: B2A1522117, E0514103, B2A15X2238, B2A1792589, A2A2242418, A2A21X2345, B2A1692413, A2A2152263, A2A22624…serial numbers: B2A1522117, E0514103, B2A15X2238, B2A1792589, A2A2242418, A2A21X2345, B2A1692413, A2A2152263, A2A2262450, A2A22Z2576, W4A1252250, A2A20Z2167, W4A10X2001, A2A2412831, W1C0552181, W1D0842501, W4D13Z2452, W4A1322350, W4D1422467, B2A1472002, A1C1982123, W2A1062134, W1B08X2042, W2A1072139, W4A1242220, A2A2382741, B2A17Z2627, A2A2152264, W4C1362387, W4D1422465, D4W1262257, W1D0892543, W4D1462478, W4A1182118, W4A11X2133, W4A11Z2167, N/A Deinstalled, A1B18Z7003, W4A1252243, DFP800A-0001, A3552017, W2A1092165, W1D0882537, A2A2122203, W4A10Z2022, W4B1332361, A2A23Y2788, W4A12Z2328, B2A1662374, W4A12Z2329, B2A1622309, B2A14X2056, A2A2272459, A2A2172291, W4A1132061, W1D0832488, W1D0882536, W2A0962022, W2A0982038, W2A0972026, B2A1642351, A2A2152255, W1D0752392, A2A2182308, A2A2412836, A2A2112189, A2A21X2336, A1C19Y2164, A1C1942056, A2A2192328, A2A2332643, W4C1372398, A2A22Y2553, A2A2132226, W4A1262256, B2A1472013, W2A1022097, W2A1022094, W4A10Z2017, W1B08Y2055, B2A1492051, B2A14920 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1179-2026%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.