RecallRadar

FDA Device recall Z-0175-2018

Dimension Vista Protein 1 Control M, PROT 1 CON M, Siemens Material Number (SMN) 10445920, Product Code KC716

On 2017-12-06 the FDA classified a Class II recall of Dimension Vista Protein 1 Control M, PROT 1 CON M, Siemens Material Number (SMN) 10445920, Product Code KC716 by Siemens Healthcare Diagnostics, citing complaint of high amount of ceruloplasmin patient results below the reference range (< 20 mg/dL [0.2 g/L]. PROT1 CAL and corresponding PROT1 CON L, M, and H are inaccurately assig….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-0175-2018
ClassificationClass II
Statusterminated
Initiated2017-05-23
Published2017-12-06
Last updated2026-09-13 11:54 UTC
CompanySiemens Healthcare Diagnostics
DistributionUS

Product

Dimension Vista Protein 1 Control M, PROT 1 CON M, Siemens Material Number (SMN) 10445920, Product Code KC716; IVD

Reason

Complaint of high amount of ceruloplasmin patient results below the reference range (< 20 mg/dL [0.2 g/L]. PROT1 CAL and corresponding PROT1 CON L, M, and H are inaccurately assigned for ceruloplasmin (CER).

Identifiers

code_info
Lot numbers: 6BQM01, 6BQM01A, 6BQM01B, 6BQM01C,6BQM01D, 6BQM01E (exp. date 2017-06-17); 6FQM01, 6FQM01A (exp. date 201…Lot numbers: 6BQM01, 6BQM01A, 6BQM01B, 6BQM01C,6BQM01D, 6BQM01E (exp. date 2017-06-17); 6FQM01, 6FQM01A (exp. date 2017-11-03) --- Lot # (UDI): 6BQM01 (008427680063336JQL01C20180131), 6BQM01A (008427680063406BQM01A20170617), 6BQM01B (008427680063406BQM01B20170617), 6BQM01C (008427680063406BQM01C20170617), 6BQM01D( 008427680063406BQM01D20170617), 6BQM01E (008427680063406BQM01E20170617), 6FQM01(008427680063406FQM0120171103), 6FQM01A (008427680063406FQM01A20171103).