RecallRadar

FDA Drug recall D-366-2014

Carbon Dioxide, USP, Medical Gas, Rx only, Non Flammable Gas 2, in 50 lbs steel cylinders, Distributed by AIRGAS USA, LLC Radnor, PA

On 2013-12-25 the FDA classified a Class II recall of Carbon Dioxide, USP, Medical Gas, Rx only, Non Flammable Gas 2, in 50 lbs steel cylinders, Distributed by AIRGAS USA, LLC Radnor, PA by Airgas Medical Services, citing labeling: Not Elsewhere Classified; Due to an error in the manufacturing procedure, a cylinder in liquid withdrawal service was not marked Syphon Tube indicating liquid withdraw….

Read the official FDA Drug notice

AgencyFDA Drug
Recall numberD-366-2014
ClassificationClass II
Statusterminated
Initiated2013-11-06
Published2013-12-25
Last updated2026-09-13 11:51 UTC
CompanyAirgas Medical Services
DistributionAR, KS, MO

Product

Carbon Dioxide, USP, Medical Gas, Rx only, Non Flammable Gas 2, in 50 lbs steel cylinders, Distributed by AIRGAS USA, LLC Radnor, PA

Reason

Labeling: Not Elsewhere Classified; Due to an error in the manufacturing procedure, a cylinder in liquid withdrawal service was not marked Syphon Tube indicating liquid withdrawal and shipped as a gas withdrawal cylinder.

Identifiers

code_infoNDC 11054-002 Lot # TY00A260A Size G and H Part number 835148

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:51 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/drug/enforcement.json?search=recall_number:%22D-366-2014%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.