RecallRadar

FDA Device recall Z-3257-2026

Brand Name: VOLARA System Product Name: VOLARA, HC, NA A Model/Catalog Number: M08594A Software Version: N/A Product Description: VOLARA, H…

On 2026-09-30 the FDA classified a Class II recall of Brand Name: VOLARA System Product Name: VOLARA, HC, NA A Model/Catalog Number: M08594A Software Version: N/A Product Description: VOLARA, H… by Baxter Healthcare, citing contamination within the release test equipment flow path could have affected the accuracy of the nebulizer flow rate. Affected devices returned to firm and testing confirmed devi….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-3257-2026
ClassificationClass II
Statusongoing
Initiated2026-08-03
Published2026-09-30
Last updated2026-10-06 05:30 UTC
CompanyBaxter Healthcare
DistributionUS

Product

Brand Name: VOLARA System Product Name: VOLARA, HC, NA A Model/Catalog Number: M08594A Software Version: N/A Product Description: VOLARA, HOMECARE, NORTH AMERICA Component: No

Reason

Contamination within the release test equipment flow path could have affected the accuracy of the nebulizer flow rate. Affected devices returned to firm and testing confirmed devices meet specifications.

Identifiers

code_infoLot Code: UDI: 00887761981690 Serial Numbers: W078SY3553 and Z309SY7853

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-10-06 05:30 UTC. Current status: ongoing.

Other recalls by Baxter Healthcare

All Baxter Healthcare recalls

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3257-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.