RecallRadar

FDA Device recall Z-3243-2018

Terumo 220/240V AC, Advanced Perfusion System Platform (APS) for use in cardiopulmonary bypass procedures

On 2018-10-03 the FDA classified a Class II recall of Terumo 220/240V AC, Advanced Perfusion System Platform (APS) for use in cardiopulmonary bypass procedures by Terumo Cardiovascular Systems, citing screws on a Terumo¿ Advanced Perfusion System 1 Power Manager Board (the base unit circuit board) were not torqued to specification. It is possible that a screw which has not been….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-3243-2018
ClassificationClass II
Statusterminated
Initiated2018-08-28
Published2018-10-03
Last updated2026-09-13 11:54 UTC
CompanyTerumo Cardiovascular Systems
DistributionUS

Product

Terumo 220/240V AC, Advanced Perfusion System Platform (APS) for use in cardiopulmonary bypass procedures. Catalog Number 801764

Reason

Screws on a Terumo¿ Advanced Perfusion System 1 Power Manager Board (the base unit circuit board) were not torqued to specification. It is possible that a screw which has not been properly torqued can go undetected during manufacturing testing and may loosen over time, resulting in the potential for an electrical short in the system

Identifiers

code_info
Serial Numbers: 1522 1523 1524 1525…Serial Numbers: 1522 1523 1524 1525 1528 1530 1526 1527 1532 1544 Spare Parts: 870509 ASSEMBLY POWER MANAGER POD APS S/N: 03097 03099 03100 03101 03113 03114 03115 03143 03155 03167 03264

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3243-2018%22