FDA Device recall Z-3237-2017

AutoCheck5+, Level 2, specifically designed for Radiometer Americas blood gas analyzers The QUALICHECK+ quality control (QC) ampoules are…

On 2017-10-11 the FDA classified a Class II recall of AutoCheck5+, Level 2, specifically designed for Radiometer Americas blood gas analyzers The QUALICHECK+ quality control (QC) ampoules are… by Radiometer America, citing radiometer has recently become aware that a few of the AUTOCHECK 5+ level 2 ampoules, will give incorrect results for pCO2 and pO2. A defective QC ampoule will for: " pCO2 give a….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-3237-2017
ClassificationClass II
Statusterminated
Initiated2017-09-07
Published2017-10-11
Last updated2026-09-13 11:54 UTC
CompanyRadiometer America
DistributionUS

Product

AutoCheck5+, Level 2, specifically designed for Radiometer Americas blood gas analyzers The QUALICHECK+ quality control (QC) ampoules are specifically designed for Radiometer America s blood gas analyzers

Reason

Radiometer has recently become aware that a few of the AUTOCHECK 5+ level 2 ampoules, will give incorrect results for pCO2 and pO2. A defective QC ampoule will for: " pCO2 give a result close to or just outside of the low end of the control range. (4.92  5.85 kPa or 36.9  46.9 mmHg) " pO2 give a considerable higher result than the expected assigned value of 13.9 -14.3 kPa (104-107 mmHg)

Identifiers

code_infoS7745 Lot R0698