FDA Device recall Z-3228-2026

Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories MDR107000P BED, MANUAL, MEDLITE

On 2026-10-07 the FDA classified a Class I recall of Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories MDR107000P BED, MANUAL, MEDLITE by Medline Industries, citing potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, spa….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-3228-2026
ClassificationClass I
Statusongoing
Initiated2026-08-14
Published2026-10-07
Last updated2026-10-13 05:30 UTC
CompanyMedline Industries
DistributionUS

Product

Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories MDR107000P BED, MANUAL, MEDLITE

Reason

Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, sparking, melting, burning, or fire. Additional products added to 11/26/2025 recall.

Identifiers

code_infoMDR107000P UDI-DI 00080196288284 All lots

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-10-13 05:30 UTC. Current status: ongoing.