RecallRadar

FDA Device recall Z-3228-2017

Platinium VR 1210, Model Number TDF031U, UDI/GTIN 08031527015514, Implantable cardioverter defibrillator, Biventricular or ventricular anti…

On 2017-10-04 the FDA classified a Class II recall of Platinium VR 1210, Model Number TDF031U, UDI/GTIN 08031527015514, Implantable cardioverter defibrillator, Biventricular or ventricular anti… by Sorin Group Italia Srl Crf, citing there is a possibility of overconsumption of certain PLATINIUM Implantable Cardiac Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillators (CRT-Ds) following an….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-3228-2017
ClassificationClass II
Statusterminated
Initiated2017-07-24
Published2017-10-04
Last updated2026-09-13 11:54 UTC
CompanySorin Group Italia Srl Crf
DistributionUS

Product

Platinium VR 1210, Model Number TDF031U, UDI/GTIN 08031527015514, Implantable cardioverter defibrillator, Biventricular or ventricular antitachycardia pacing, Dual or single chamber arrhythmia detection Product Usage: PLATINIUM VR is indicated for use in patients who are at high risk of sudden cardiac death due to ventricular tachyarrhythmias and who have experienced one of the following situations: - Survival of at least one episode of cardiac arrest (manifested by the loss of consciousness) due to ventricular tachyarrhythmia, - Recurrent, poorly tolerated sustained ventricular tachycardia (VT).

Reason

There is a possibility of overconsumption of certain PLATINIUM Implantable Cardiac Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillators (CRT-Ds) following an ElectroStatic Discharge (ESD) during the implant surgery or a Magnetic Resonance Imaging (MRI) scan, and what actions to take.

Identifiers

code_infoAll Serial Numbers

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3228-2017%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.