RecallRadar

FDA Device recall Z-3204-2026

TRIDENT 0¿ POLY INSERT 28mm ID 620-00-28D Orthopaedic implant components manufactured from UHMWPE (polyethylene)

On 2026-09-23 the FDA classified a Class II recall of TRIDENT 0¿ POLY INSERT 28mm ID 620-00-28D Orthopaedic implant components manufactured from UHMWPE (polyethylene) by Howmedica Osteonics, citing affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-3204-2026
ClassificationClass II
Statusongoing
Initiated2026-07-30
Published2026-09-23
Last updated2026-09-29 05:30 UTC
CompanyHowmedica Osteonics
DistributionUS

Product

TRIDENT 0¿ POLY INSERT 28mm ID 620-00-28D Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Reason

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

Identifiers

code_info620-00-28D UDI-DI 07613327313581 Lots MHANP2 MHD1N0

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-29 05:30 UTC. Current status: ongoing.

Other recalls by Howmedica Osteonics

All Howmedica Osteonics recalls

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3204-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.