FDA Device recall Z-3201-2026
ACETABULAR CUP UHMWPE 61-3258 ALL POLY CONSTRAINED INSERT 44MM 69-2244 ALL POLY CONSTRAINED INSERT 46MM 69-2246 ALL POLY CONSTRAINED INSERT…
- FDA Device
- Class II
- ongoing
- Published 2026-09-23
On 2026-09-23 the FDA classified a Class II recall of ACETABULAR CUP UHMWPE 61-3258 ALL POLY CONSTRAINED INSERT 44MM 69-2244 ALL POLY CONSTRAINED INSERT 46MM 69-2246 ALL POLY CONSTRAINED INSERT… by Howmedica Osteonics, citing affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-3201-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2026-07-30 |
| Published | 2026-09-23 |
| Last updated | 2026-09-29 05:30 UTC |
| Company | Howmedica Osteonics |
| Distribution | US |
Product
ACETABULAR CUP UHMWPE 61-3258 ALL POLY CONSTRAINED INSERT 44MM 69-2244 ALL POLY CONSTRAINED INSERT 46MM 69-2246 ALL POLY CONSTRAINED INSERT 48MM 69-2248 ALL POLY CONSTRAINED INSERT 50MM 69-2850 ALL POLY CONSTRAINED INSERT 52MM 69-2852 CONSTRAINED ACETABULAR INSERTS 22MM 50MM 2099-2250 CONSTRAINED ACETABULAR INSERTS 22MM 52MM 2099-2252 CONSTRAINED ACETABULAR INSERTS 22MM 56MM 2099-2256 CONSTRAINED ACETABULAR INSERTS 28MM 61MM 2099-2861 EIUS UNI TIB SMALL RM/LL 9MM 6636-0-419 PCA CONST LINER 32MM NEUTRAL C - SHELL 6299-N-332 SYS 12 CONST LINER 32ID 10 DEG P6 +4MM 6302-H-632 TRIDENT 0 DEG CONSTRAINED INSERT 22D 690-00-22D TRIDENT 0 DEG CONSTRAINED INSERT 22E 690-00-22E TRIDENT 0 DEG CONSTRAINED INSERT 28F 690-00-28F TRIDENT 0 DEG CONSTRAINED INSERT 28G 690-00-28G TRIDENT 0 DEG CONSTRAINED INSERT 32I 690-00-32I TRIDENT 0¿ POLY INSERT 28mm ID 620-00-28D TRIDENT CONSTR ACET INSERT 690-10-28K TRIDENT CONSTRAINED INSERT 690-10-22E 690-10-22F 690-10-28G 690-10-28H 690-10-28I Orthopaedic implant components manufactured from UHMWPE (polyethylene)
Reason
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Identifiers
| code_info | 61-3258 UDI-DI 07613327038538 Lots 191723 191867 69-2244 UDI-DI 07613327025446 Lots 990V1W D56JV4…61-3258 UDI-DI 07613327038538 Lots 191723 191867 69-2244 UDI-DI 07613327025446 Lots 990V1W D56JV4 D74X5E DX56K2 E67D6N EP10WL KP14HN NN8NM2 P36TXM P95L8X RE4EK0 X82K0D 69-2246 UDI-DI 07613327025453 Lots 8X6EL1 69-2248 UDI-DI 07613327025460 Lots 9J3DPH A41E7D AD39A3 RP673W 69-2850 UDI-DI 07613327025477 Lots MLL1M3 MLLXT1 MLLY6H 69-2852 UDI-DI 07613327025484 Lots MLL0NM MLLAHY 2099-2250 UDI-DI 07613327031430 Lots 1K5VVW 2367AK 3X0NJD 4W4MPL LV80NE 2099-2252 UDI-DI 07613327031447 Lots NV561W PM326V 2099-2256 UDI-DI 07613327031454 Lots 3H5RRK AA7JK0 2099-2861 UDI-DI 07613327031591 Lot MLLVL6 6636-0-419 No UDI Lot 13SFLA 6299-N-332 UDI-DI 07613327138870 Lot 19ECK 6302-H-632 UDI-DI 07613327138979 Lot 15NKT 690-00-22D UDI-DI 07613327025538 Lots 123J5R 194RN0 274HMK 3M77NN 5J8P89 5L82T0 986HVM A23P1X E923WW M |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-29 05:30 UTC. Current status: ongoing.
Other recalls by Howmedica Osteonics
- BEARING AND AXLE BUSHING ASSY STD ELBOW C-M113-1-301 KRH REV TIB INS 16mm THK .640 DIA STEM C-M113-5-600 SELF-PRESSURIZING GLENOID SIZE 9 C…
FDA Device · Class II · 2026-09-23 · Howmedica Osteonics - TRIATHLON HINGE INSERT SIZE 4 11MM 5612-P-411 TRIATHLON HINGE INSERT SIZE 4 13MM 5612-P-413 Orthopaedic implant components manufactured fro…
FDA Device · Class II · 2026-09-23 · Howmedica Osteonics - SCORPIO C-DOME PATELLA 73-2710 73-2910 SCORPIO U-DOME PATELLA 73-3110 73-3508 73-3708 73-3710 73-3910 Orthopaedic implant components manufa…
FDA Device · Class II · 2026-09-23 · Howmedica Osteonics - DURACON XL ANGLED TIB INS 11/14MM C-M064-1-300 SELF-PRESSURIZING GLENOID SIZE 8 C-M062-6-800 C-M062-7-500 C-M063-2-600 SELF-PRESSURIZING GL…
FDA Device · Class II · 2026-09-23 · Howmedica Osteonics - AXLE PIN ASSY LARGE ELBOW COBALT CHROME Catalog No. C-M112-2-500 Orthopaedic implant components manufactured from UHMWPE (polyethylene)
FDA Device · Class II · 2026-09-23 · Howmedica Osteonics
All Howmedica Osteonics recalls
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3201-2026%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.