RecallRadar

FDA Device recall Z-3200-2024

Ingenia Elition X- For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal str…

On 2024-10-02 the FDA classified a Class II recall of Ingenia Elition X- For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal str… by Philips North America, citing integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially bl….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-3200-2024
ClassificationClass II
Statusongoing
Initiated2024-09-03
Published2024-10-02
Last updated2026-09-13 11:57 UTC
CompanyPhilips North America
DistributionUS

Product

Ingenia Elition X- For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 1) 782151; 2) 782119; 3) 781358; 4) 782107; 5) 782136

Reason

Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.

Identifiers

code_info
Model UDI-DI: 782151 N/A; 782119 (01)00884838104129(21); 781358 (01)00884838088115(21); 782107 (01)0088…Model UDI-DI: 782151 N/A; 782119 (01)00884838104129(21); 781358 (01)00884838088115(21); 782107 (01)00884838098336(21); 782136 (01)00884838108608(21) Serial Numbers: 26500 26505 28502 28509 45393 45467 45599 62029 62041 62051 62052 62061 62066 62070 62071 62081 62082 62083 62091 62103 62111 62113 62114 62120 62126 62129 62147 62148 62186 62187 62188 66001 66002

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3200-2024%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.