FDA Device recall Z-3199-2026
ABG I ACETABULAR INSERT HOODED 28mm ID X 48mm C-M079-2-801 ARC2F ACETABULAR INSERT 48/50/52mm OD x 28mm ID C-M071-6-700 ARC2F ACETABULAR IN…
- FDA Device
- Class II
- ongoing
- Published 2026-09-23
On 2026-09-23 the FDA classified a Class II recall of ABG I ACETABULAR INSERT HOODED 28mm ID X 48mm C-M079-2-801 ARC2F ACETABULAR INSERT 48/50/52mm OD x 28mm ID C-M071-6-700 ARC2F ACETABULAR IN… by Howmedica Osteonics, citing affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-3199-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2026-07-30 |
| Published | 2026-09-23 |
| Last updated | 2026-09-29 05:30 UTC |
| Company | Howmedica Osteonics |
| Distribution | US |
Product
ABG I ACETABULAR INSERT HOODED 28mm ID X 48mm C-M079-2-801 ARC2F ACETABULAR INSERT 48/50/52mm OD x 28mm ID C-M071-6-700 ARC2F ACETABULAR INSERT 50mm OD x 28mm ID 10DEGREE HOOD C-M071-8-400 BEARING AND AXLE BUSHING ASSY STANDARD ELBOW C-M110-1-201 BUSHING AND BEARING USE C-MM100253-00 C-MM100253-02 DURACON TIB INSERT LG 12MM MED 14MM LAT C-M066-3-000 EXETER WINGLESS CENTRALISER FOR C-M082-7-302 C-M082-7-303 EXETER WINGLESS CENTRALIZER XSM C-M107-5-301 GMRS EXTRA SM ALL-POLY TIBIA SIZE 11mm C-MM100616-02 GMRS EXTRA SM ALL-POLY TIBIA SIZE 16mm C-MM100616-03 GMRS EXTRA SM ALL-POLY TIBIA SIZE 8mm C-MM100616-01 GMRS SML POLY TIBIA 11MM CASE 100520 C-MM100520-03 GMRS SML POLY TIBIA 16MM CASE 100520 C-MM100520-04 GMRS SML POLY TIBIA 8MM CASE 100520 C-MM100520-02 GMRS SML TIBIA 11MM WITH CASE 100411 C-MM100411-03 GMRS SML TIBIA 14MM WITH CASE 100411 C-MM100411-04 GMRS SML TIBIA 8MM WITH CASE 100411 C-MM100411-02 GMRS TIB INSRT SML 10MM WITH CASE 100507 C-MM100507-04 GMRS TIB INSRT SML 16MM WITH CASE 100507 C-MM100507-06 KALI TIBIAL INSERT LARGE RIGHT 8mm C-M071-6-601 KRH REV TIB INS 16mm THK .655 DIA STEM C-M113-5-603 MRH KNEE TIBIAL INSERT FOR THIN POST C-M067-3-103 MRS RADIAL BUSHING FORITH C-M107-7-800 C-M107-7-801 SELF-PRESSURIZING GLENOID SIZE 9 C-M064-4-300 TRIATHLON NO. 9 TIBIAL INSERT 11MM C-M067-5-003 TRIATHLON NO. 9 TIBIAL INSERT 9MM C-M067-5-002 TRIDENT ALL POLY CUP 44 X 66MM C-M066-0-901 TRIDENT CONSTRAINED ACETABULAR INSERT SIZE A X 22mm ID UH1 C-M104-2-002 TRIDENT CONSTRAINED ACETABULAR INSERT SIZE B X 22mm ID UH1 C-M104-2-003 UHR BIPOLAR 22X38MM UH1-38-22 UHR BIPOLAR 22X40MM UH1-40-22 UHR BIPOLAR 22X41MM UH1-41-22 UHR BIPOLAR 22X42MM UH1-42-22 UHR BIPOLAR 22X43MM UH1-43-22 UHR BIPOLAR 22X45MM UH1-45-22 UHR BIPOLAR 26X41mm UH1-41-26 UHR BIPOLAR 26X42mm UH1-42-26 UHR BIPOLAR 26X43mm UH1-43-26 UHR BIPOLAR 26X44mm UH1-44-26 UHR BIPOLAR 26X45mm UH1-45-26 UHR BIPOLAR 26X46mm UH1-46-26 UHR BIPOLAR 26X47mm UH1-47-26 UHR BIPOLAR 26X48mm UH1-48-26 UHR BIPOLAR 28X42mm UH1-42-28 UHR BIPOLAR 28X44mm UH1-44-28 UHR BIPOLAR 28X45mm UH1-45-28 UHR BIPOLAR 28X46mm UH1-46-28 UHR BIPOLAR 28X47mm UH1-47-28 UHR BIPOLAR 28X48mm UH1-48-28 Orthopaedic implant components manufactured from UHMWPE (polyethylene)
Reason
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Identifiers
| code_info | C-M079-2-801 No UDI Lot G3264034 C-M071-6-700 No UDI Lot G3043883 C-M071-8-400 No UDI Lot G304387…C-M079-2-801 No UDI Lot G3264034 C-M071-6-700 No UDI Lot G3043883 C-M071-8-400 No UDI Lot G3043878 C-M110-1-201 UDI-DI 07613327504811 Lot 573P37 C-MM100253-02 No UDI Lot N8873H C-M066-3-000 No UDI Lot RX06630A C-M082-7-303 No UDI Lot L6237 C-M107-5-301 No UDI Lot 193607 C-MM100616-02 No UDI Lot 195261 C-MM100616-03 No UDI Lot 195262 C-MM100616-01 No UDI Lot 195260 C-MM100520-03 No UDI Lot 195081 C-MM100520-04 No UDI Lot 195082 C-MM100520-02 No UDI Lot 195080 C-MM100411-03 No UDI Lot 194962 C-MM100411-04 No UDI Lot 194963 C-MM100411-02 No UDI Lot 194961 C-MM100507-04 No UDI Lot LNH795 C-MM100507-06 No UDI Lot LNH797 C-M071-6-601 No UDI Lot G3055684 C-M113-5-603 No UDI Lot 194594 C-M067-3-103 No UDI Lot MHP0AX C-M107-7-801 No UDI Lot 193479 C-M064-4-300 No UDI Lot RX06443A C-M067-5-003 No UDI |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-29 05:30 UTC. Current status: ongoing.
Other recalls by Howmedica Osteonics
- BEARING AND AXLE BUSHING ASSY STD ELBOW C-M113-1-301 KRH REV TIB INS 16mm THK .640 DIA STEM C-M113-5-600 SELF-PRESSURIZING GLENOID SIZE 9 C…
FDA Device · Class II · 2026-09-23 · Howmedica Osteonics - TRIATHLON HINGE INSERT SIZE 4 11MM 5612-P-411 TRIATHLON HINGE INSERT SIZE 4 13MM 5612-P-413 Orthopaedic implant components manufactured fro…
FDA Device · Class II · 2026-09-23 · Howmedica Osteonics - SCORPIO C-DOME PATELLA 73-2710 73-2910 SCORPIO U-DOME PATELLA 73-3110 73-3508 73-3708 73-3710 73-3910 Orthopaedic implant components manufa…
FDA Device · Class II · 2026-09-23 · Howmedica Osteonics - DURACON XL ANGLED TIB INS 11/14MM C-M064-1-300 SELF-PRESSURIZING GLENOID SIZE 8 C-M062-6-800 C-M062-7-500 C-M063-2-600 SELF-PRESSURIZING GL…
FDA Device · Class II · 2026-09-23 · Howmedica Osteonics - AXLE PIN ASSY LARGE ELBOW COBALT CHROME Catalog No. C-M112-2-500 Orthopaedic implant components manufactured from UHMWPE (polyethylene)
FDA Device · Class II · 2026-09-23 · Howmedica Osteonics
All Howmedica Osteonics recalls
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3199-2026%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.