FDA Device recall Z-3199-2024
Ingenia Elition S -For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal str…
- FDA Device
- Class II
- ongoing
- Published 2024-10-02
On 2024-10-02 the FDA classified a Class II recall of Ingenia Elition S -For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal str… by Philips North America, citing integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially bl….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-3199-2024 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2024-09-03 |
| Published | 2024-10-02 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Philips North America |
| Distribution | US |
Product
Ingenia Elition S -For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 1) 782150 2) 782106
Reason
Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.
Identifiers
| code_info | Model UDI-DI: 782150 (01)00884838098329(21); 782106 (01)00884838098329(21) Serial Numbers: 26020…Model UDI-DI: 782150 (01)00884838098329(21); 782106 (01)00884838098329(21) Serial Numbers: 26020 26023 26026 26034 46240 46270 46310 46315 46324 46352 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3199-2024%22