FDA Device recall Z-3196-2024
Ingenia 3.0T CX - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal stru…
- FDA Device
- Class II
- ongoing
- Published 2024-10-02
On 2024-10-02 the FDA classified a Class II recall of Ingenia 3.0T CX - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal stru… by Philips North America, citing integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially bl….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-3196-2024 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2024-09-03 |
| Published | 2024-10-02 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Philips North America |
| Distribution | US |
Product
Ingenia 3.0T CX - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 782105
Reason
Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.
Identifiers
| code_info | UDI-DI: (01)00884838098312(21) Serial Numbers: 78547 78557 78559 78560 78562 78564 78565 78566 |
|---|