RecallRadar

FDA Device recall Z-3192-2026

Et Control, Software option for anesthesia gas machine to achieve and maintain targeted end tidal oxygen and anesthetic agents, Field repla…

On 2026-09-23 the FDA classified a Class II recall of Et Control, Software option for anesthesia gas machine to achieve and maintain targeted end tidal oxygen and anesthetic agents, Field repla… by Datex Ohmeda, citing anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 1….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-3192-2026
ClassificationClass II
Statusongoing
Initiated2026-07-30
Published2026-09-23
Last updated2026-09-29 05:30 UTC
CompanyDatex Ohmeda
DistributionUS

Product

Et Control, Software option for anesthesia gas machine to achieve and maintain targeted end tidal oxygen and anesthetic agents, Field replaceable unit (FRU) part numbers: 1011-8000-000-S, 5700665

Reason

Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.

Identifiers

code_info
Serial Numbers: 801042711 801031045 801031926 801045423 801011743 801090568 801128677 801011276-1 801011276-10 80101127…Serial Numbers: 801042711 801031045 801031926 801045423 801011743 801090568 801128677 801011276-1 801011276-10 801011276-11 801011276-12 801011276-13 801011276-14 801011276-15 801011276-2 801011276-3 801011276-4 801011276-5 801011276-6 801011276-7 801011276-8 801011276-9 801078342-1 801078342-10 801078342-11 801078342-12 801078342-13 801078342-14 801078342-15 801078342-16 801078342-17 801078342-18 801078342-19 801078342-2 801078342-20 801078342-21 801078342-22 801078342-23 801078342-24 801078342-25 801078342-26 801078342-27 801078342-28 801078342-29 801078342-3 801078342-30 801078342-4 801078342-5 801078342-6 801078342-7 801078342-8 801078342-9 801102924 801108227 801113705 801114965 801084897 801003006 801004394 801071735 801016238 801050584 801051585 801064585 801013238 801013095 801043259 801088072 801014199 801006762 801052860 801023283 801110573 801064058 801080036 801126195 801064184-1 801064184-2 801059320-1 801059320-2 801091614 801028252 801051344 801030002 800988180 801111226

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-29 05:30 UTC. Current status: ongoing.

Other recalls by Datex Ohmeda

All Datex Ohmeda recalls

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3192-2026%22