Xper Flex Cardio Physiomonitoring System Model # 453564669081 Software Version 1.6.0.0388

On 2018-09-26 the FDA classified a Class II recall of Xper Flex Cardio Physiomonitoring System Model # 453564669081 Software Version 1.6.0.0388 by Invivo, citing released with an incorrect, older firmware version installed. May result in intermittent loss of ECG monitoring, inaccurate ECG amplitude display and or auxiliary output, and inac….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-3192-2018
ClassificationClass II
Statusterminated
Initiated2018-08-03
Published2018-09-26
Last updated2026-09-13 11:54 UTC
CompanyInvivo
Distributionn/a

Product

Xper Flex Cardio Physiomonitoring System Model # 453564669081 Software Version 1.6.0.0388

Reason

Released with an incorrect, older firmware version installed. May result in intermittent loss of ECG monitoring, inaccurate ECG amplitude display and or auxiliary output, and inaccurate heart rate (HR) displayed due to QRS detection fault.

Identifiers

code_info
System Serial Numbers: US1814000001, US1814000002, US1814000003, US1814000004, US1814000006, US1814000007, US181400000…System Serial Numbers: US1814000001, US1814000002, US1814000003, US1814000004, US1814000006, US1814000007, US1814000008, US1814000011, US1814000012, US1814000013, US1814000014, US1814000016, US1814000017, US1814000018, US1814000019, US1814000020, US1814000022, US1814000023, US1814000024, US1814000027, US1814000028, US1814000030, US1814000031, US1814000032, US1814000033, US1814000034, US1814000037, US1814000044, US1814000046, US1814000050, US1814000053, US1814000062, US1814000066, US1814000068,and US1814000072