FDA Device recall Z-3175-2024
STREAMLINE FRESENIUS FOR DAVITA-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis
- FDA Device
- Class II
- ongoing
- Published 2024-09-25
On 2024-09-25 the FDA classified a Class II recall of STREAMLINE FRESENIUS FOR DAVITA-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis by B Braun Medical, citing incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rin….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-3175-2024 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2024-08-15 |
| Published | 2024-09-25 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | B Braun Medical |
| Distribution | US |
Product
STREAMLINE FRESENIUS FOR DAVITA-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2000M2095DA
Reason
Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method
Identifiers
| code_info | UDI-DI 04046955776023; All Unexpired Lot Numbers |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3175-2024%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.