FDA Device recall Z-3170-2024
ADVIA Chemistry Microalbumin_2 (¿ALB_2) -For in vitro diagnostic use in the quantitative measurement of microalbumin in human urine on ADVI…
- FDA Device
- Class II
- ongoing
- Published 2024-09-25
On 2024-09-25 the FDA classified a Class II recall of ADVIA Chemistry Microalbumin_2 (¿ALB_2) -For in vitro diagnostic use in the quantitative measurement of microalbumin in human urine on ADVI… by Siemens Healthcare Diagnostics, citing atellica CH Microalbumin_2 (¿ALB_2) and the ADVIA Chemistry Microalbumin_2 (¿ALB_2) are not meeting the High-Dose Hook Effect claim/Prozone Effect claim as stated in the Instructi….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-3170-2024 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2024-08-15 |
| Published | 2024-09-25 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Siemens Healthcare Diagnostics |
| Distribution | US |
Product
ADVIA Chemistry Microalbumin_2 (¿ALB_2) -For in vitro diagnostic use in the quantitative measurement of microalbumin in human urine on ADVIA Chemistry systems/ Atellica CH and Atellica CI Analyzers.
Reason
Atellica CH Microalbumin_2 (¿ALB_2) and the ADVIA Chemistry Microalbumin_2 (¿ALB_2) are not meeting the High-Dose Hook Effect claim/Prozone Effect claim as stated in the Instructions for Use (IFU) on the Atellica CH, Atellica CI Analyzers, ADVIA 1800 Chemistry Systems, ADVIA 2400 Chemistry Systems, and ADVIA Chemistry XPT Systems, may result in erroneously depressed microalbumin patient results.
Identifiers
| code_info | UDI-DI: 00630414517643 Lot Numbers: 612528, 616644, 620971, 623768, 636498, 640513, 658356, 668872 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3170-2024%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.