RecallRadar

FDA Device recall Z-3160-2024

GE Healthcare Centricity Universal Viewer Zero Footprint Client, Model Numbers: a) 2110344-008

On 2024-09-25 the FDA classified a Class II recall of GE Healthcare Centricity Universal Viewer Zero Footprint Client, Model Numbers: a) 2110344-008 by Ge Healthcare, citing footprint Client (ZFP) versions v6.0 SP9.x and SP10.x where the latest addended report is not shown by default to the user. The issue occurs when the addendum is created on the sa….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-3160-2024
ClassificationClass II
Statusongoing
Initiated2024-07-23
Published2024-09-25
Last updated2026-09-13 11:57 UTC
CompanyGe Healthcare
Distributionn/a

Product

GE Healthcare Centricity Universal Viewer Zero Footprint Client, Model Numbers: a) 2110344-008; b) 2110344-009; c) 2110344-010; d) 2110344-011; e) 2110344-013; f) 2110344-019; g) 2110344-025; h) 2110344-029; i) 2110344-030; j) 2110344-032 ; Radiological image processing system

Reason

Footprint Client (ZFP) versions v6.0 SP9.x and SP10.x where the latest addended report is not shown by default to the user. The issue occurs when the addendum is created on the same day as the original report but at a different time, and only when launched on the following browsers. " Internet Explorer 10 (IE10) " Internet Explorer 11 (IE11) " Firefox.

Identifiers

code_info
UDI/DI: 00840682102988: a) 2110344-008, Installed Product ID Numbers: B54834-Centricity Universal Viewer Zero Footp…UDI/DI: 00840682102988: a) 2110344-008, Installed Product ID Numbers: B54834-Centricity Universal Viewer Zero Footprint / ZFP-02394744, KMCH-Centricity Universal Viewer Zero Footprint / ZFP-01819848, 355920-Centricity Universal Viewer Zero Footprint / ZFP-20996307, 225717-Centricity Universal Viewer Zero Footprint / ZFP-19666806, RF0827-Centricity Universal Viewer Zero Footprint / ZFP-19460712; b) 2110344-009, Installed Product ID Numbers: 1472-15-Centricity Universal Viewer Zero Footprint / ZFP-01372918, 1495-3-Centricity Universal Viewer Zero Footprint / ZFP-02348000, 1495-3-Centricity Universal Viewer Zero Footprint / ZFP-02348009, 1495-3-Centricity Universal Viewer Zero Footprint / ZFP-02348010, 1495-3-Centricity Universal Viewer Zero Footprint / ZFP-02348035, 1407636-Centricity Universal Viewer Zero Footprint / ZFP-19632983, SA1008-Centricity Universal Viewer Zero Footprint / ZFP-20620864; c) 2110344-010, Installed Product ID Numbers: 3922-18-Centricity Universal Viewer Ze

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3160-2024%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.