FDA Device recall Z-3160-2024
GE Healthcare Centricity Universal Viewer Zero Footprint Client, Model Numbers: a) 2110344-008
- FDA Device
- Class II
- ongoing
- Published 2024-09-25
On 2024-09-25 the FDA classified a Class II recall of GE Healthcare Centricity Universal Viewer Zero Footprint Client, Model Numbers: a) 2110344-008 by Ge Healthcare, citing footprint Client (ZFP) versions v6.0 SP9.x and SP10.x where the latest addended report is not shown by default to the user. The issue occurs when the addendum is created on the sa….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-3160-2024 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2024-07-23 |
| Published | 2024-09-25 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Ge Healthcare |
| Distribution | n/a |
Product
GE Healthcare Centricity Universal Viewer Zero Footprint Client, Model Numbers: a) 2110344-008; b) 2110344-009; c) 2110344-010; d) 2110344-011; e) 2110344-013; f) 2110344-019; g) 2110344-025; h) 2110344-029; i) 2110344-030; j) 2110344-032 ; Radiological image processing system
Reason
Footprint Client (ZFP) versions v6.0 SP9.x and SP10.x where the latest addended report is not shown by default to the user. The issue occurs when the addendum is created on the same day as the original report but at a different time, and only when launched on the following browsers. " Internet Explorer 10 (IE10) " Internet Explorer 11 (IE11) " Firefox.
Identifiers
| code_info | UDI/DI: 00840682102988: a) 2110344-008, Installed Product ID Numbers: B54834-Centricity Universal Viewer Zero Footp…UDI/DI: 00840682102988: a) 2110344-008, Installed Product ID Numbers: B54834-Centricity Universal Viewer Zero Footprint / ZFP-02394744, KMCH-Centricity Universal Viewer Zero Footprint / ZFP-01819848, 355920-Centricity Universal Viewer Zero Footprint / ZFP-20996307, 225717-Centricity Universal Viewer Zero Footprint / ZFP-19666806, RF0827-Centricity Universal Viewer Zero Footprint / ZFP-19460712; b) 2110344-009, Installed Product ID Numbers: 1472-15-Centricity Universal Viewer Zero Footprint / ZFP-01372918, 1495-3-Centricity Universal Viewer Zero Footprint / ZFP-02348000, 1495-3-Centricity Universal Viewer Zero Footprint / ZFP-02348009, 1495-3-Centricity Universal Viewer Zero Footprint / ZFP-02348010, 1495-3-Centricity Universal Viewer Zero Footprint / ZFP-02348035, 1407636-Centricity Universal Viewer Zero Footprint / ZFP-19632983, SA1008-Centricity Universal Viewer Zero Footprint / ZFP-20620864; c) 2110344-010, Installed Product ID Numbers: 3922-18-Centricity Universal Viewer Ze |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3160-2024%22