FDA Device recall Z-3158-2024

The device is a mobile x-ray system designed to work with Fujifilm's GOS and CsI scintillator FDR D-EVO2 (DR-ID) 12XXSE) family of digital…

On 2024-10-02 the FDA classified a Class II recall of The device is a mobile x-ray system designed to work with Fujifilm's GOS and CsI scintillator FDR D-EVO2 (DR-ID) 12XXSE) family of digital… by Fujifilm Healthcare Americas, citing when the key switch is turned on, the exposure conditions are set to the default values (85kV, 1.6 mAs, approx. 0.13mGy@ SID 1m) intended for adult chest imaging. In addition, if….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-3158-2024
ClassificationClass II
Statusongoing
Initiated2023-11-23
Published2024-10-02
Last updated2026-09-13 11:57 UTC
CompanyFujifilm Healthcare Americas
DistributionUS

Product

The device is a mobile x-ray system designed to work with Fujifilm's GOS and CsI scintillator FDR D-EVO2 (DR-ID) 12XXSE) family of digital x-ray detectors coupled.

Reason

When the key switch is turned on, the exposure conditions are set to the default values (85kV, 1.6 mAs, approx. 0.13mGy@ SID 1m) intended for adult chest imaging. In addition, if exposure conditions are set from the console while the key switch is off, the set exposure conditions will not be reflected.

Identifiers

code_infoAll serial numbers shipped from FUJIFILM

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.