FDA Device recall Z-3132-2017

FFR Link-FFR Signal Processing Module, Material Number H7495551000 It is intended to condition physiological signals from measuring devices…

On 2017-09-20 the FDA classified a Class II recall of FFR Link-FFR Signal Processing Module, Material Number H7495551000 It is intended to condition physiological signals from measuring devices… by Boston Scientific, citing the device history record (DHR) was missing its test documentation for final HIPOT (high potential) electrical testing.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-3132-2017
ClassificationClass II
Statusterminated
Initiated2017-05-15
Published2017-09-20
Last updated2026-09-13 11:54 UTC
CompanyBoston Scientific
DistributionMO

Product

FFR Link-FFR Signal Processing Module, Material Number H7495551000 It is intended to condition physiological signals from measuring devices (BSC Pressure Guidewire or an external pressure transducer), transmit and receive via radiofrequency, and recondition the signals so they can be displayed on and/or recorded in a receiving device (iLab POLARIS Multi-Modality Guidance System or other monitoring device). The physiological signals can also be distributed by cable

Reason

The device history record (DHR) was missing its test documentation for final HIPOT (high potential) electrical testing.

Identifiers

code_infoLot/Batch No. (Exp Date): SPM01975 (09/05/2044), SPM00890 (09/27/2043), SPM01616 (06/28/2044)