RecallRadar

FDA Device recall Z-3126-2024

Brand Name: PROBEAT Product Name: PROBEAT-CR Model/Catalog Number: N/A Software Version: N/A Product Description: The PROBEAT-CR is a proto…

On 2024-09-18 the FDA classified a Class II recall of Brand Name: PROBEAT Product Name: PROBEAT-CR Model/Catalog Number: N/A Software Version: N/A Product Description: The PROBEAT-CR is a proto… by Hitachi America Particle Therapy Division, citing during positioning within the system's 3D to 3D matching mode while transferring CBCT images, when the Float Image Selection button is pressed, it may display CBCT images that are….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-3126-2024
ClassificationClass II
Statusongoing
Initiated2024-05-27
Published2024-09-18
Last updated2026-09-13 11:57 UTC
CompanyHitachi America Particle Therapy Division
DistributionUS

Product

Brand Name: PROBEAT Product Name: PROBEAT-CR Model/Catalog Number: N/A Software Version: N/A Product Description: The PROBEAT-CR is a proton beam irradiation system, which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam with the prescribed dose, dose distribution and directed to the prescribed patient treatment site. Component: N/A

Reason

During positioning within the system's 3D to 3D matching mode while transferring CBCT images, when the Float Image Selection button is pressed, it may display CBCT images that are insufficient in number.

Identifiers

code_infoLot Code: (01)04560333350006(21)CRUS01(11)190920

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3126-2024%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.