FDA Device recall Z-3110-2026
Product Name: DermaV Laser System Model/Catalog Number: DermaV Medical laser device that delivers pulsed laser energy at wavelengths of 532…
- FDA Device
- Class II
- ongoing
- Published 2026-09-16
On 2026-09-16 the FDA classified a Class II recall of Product Name: DermaV Laser System Model/Catalog Number: DermaV Medical laser device that delivers pulsed laser energy at wavelengths of 532… by Lutronic, citing utilizing higher fluences at 1064nm wavelength with current cryogen parameters may result in a spark/loud noise during treatment, potentially resulting in patient burn.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-3110-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2026-08-06 |
| Published | 2026-09-16 |
| Last updated | 2026-09-23 05:30 UTC |
| Company | Lutronic |
| Distribution | US |
Product
Product Name: DermaV Laser System Model/Catalog Number: DermaV Medical laser device that delivers pulsed laser energy at wavelengths of 532 nm and 1064 nm through a fiber-optic handpiece. The system may be configured with an Intelligent Cooling Device (ICD) to provide epidermal cooling during laser treatment
Reason
Utilizing higher fluences at 1064nm wavelength with current cryogen parameters may result in a spark/loud noise during treatment, potentially resulting in patient burn.
Identifiers
| code_info | Lot Code: Model: DermaV Software Version: 1.06 UDI #: 08800291820200 Serial Numbers: DV025A010, DV025A011, DV025A012, D…Lot Code: Model: DermaV Software Version: 1.06 UDI #: 08800291820200 Serial Numbers: DV025A010, DV025A011, DV025A012, DV025A013, DV025A014, DV025A015 DV025A016, DV025A018, DV025A019, DV025A020, DV025A021, DV025A022, DV025B011, DV025B012, DV025B013, DV025B014, DV025B015, DV025D006, DV025D007, DV025D008, DV025D009, DV025D010, DV025E003, DV025E004, DV025E005, DV025E006, DV025E007, DV025E008, DV025E009, DV025E010, DV025E011, DV025E012, DV025E013, DV025E014, DV025E016, DV025E017, DV025F012, DV025F013, DV025F014, DV025F015, DV025F016, DV025G011, DV025G012, DV025G013, DV025G014, DV025G015, DV025G016, DV025G017, DV025G018, DV025G019, DV025G020, DV025J021, DV025J022, DV025J023, DV025J025, DV025J026, DV025J027, DV025J028, DV025J034, DV025J035, DV025J036, DV025J037, DV025J038, DV025J040, DV025K018, DV025K019, DV025K020, DV025K021, DV025K023, DV025K024, DV025K025, DV025K027, DV025M031, DV025M032, DV025M033, DV025M034, DV026B001, DV026B002, DV026B003, DV026B004, DV026B005, DV026B010, DV026B014, DV |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-23 05:30 UTC. Current status: ongoing.
Other recalls by Lutronic
- CLARITY II Laser System
FDA Device · Class II · 2026-03-18 · Lutronic - XERF EFFECTOR 60
FDA Device · Class II · 2025-10-22 · Lutronic
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3110-2026%22