RecallRadar

FDA Device recall Z-3110-2026

Product Name: DermaV Laser System Model/Catalog Number: DermaV Medical laser device that delivers pulsed laser energy at wavelengths of 532…

On 2026-09-16 the FDA classified a Class II recall of Product Name: DermaV Laser System Model/Catalog Number: DermaV Medical laser device that delivers pulsed laser energy at wavelengths of 532… by Lutronic, citing utilizing higher fluences at 1064nm wavelength with current cryogen parameters may result in a spark/loud noise during treatment, potentially resulting in patient burn.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-3110-2026
ClassificationClass II
Statusongoing
Initiated2026-08-06
Published2026-09-16
Last updated2026-09-23 05:30 UTC
CompanyLutronic
DistributionUS

Product

Product Name: DermaV Laser System Model/Catalog Number: DermaV Medical laser device that delivers pulsed laser energy at wavelengths of 532 nm and 1064 nm through a fiber-optic handpiece. The system may be configured with an Intelligent Cooling Device (ICD) to provide epidermal cooling during laser treatment

Reason

Utilizing higher fluences at 1064nm wavelength with current cryogen parameters may result in a spark/loud noise during treatment, potentially resulting in patient burn.

Identifiers

code_info
Lot Code: Model: DermaV Software Version: 1.06 UDI #: 08800291820200 Serial Numbers: DV025A010, DV025A011, DV025A012, D…Lot Code: Model: DermaV Software Version: 1.06 UDI #: 08800291820200 Serial Numbers: DV025A010, DV025A011, DV025A012, DV025A013, DV025A014, DV025A015 DV025A016, DV025A018, DV025A019, DV025A020, DV025A021, DV025A022, DV025B011, DV025B012, DV025B013, DV025B014, DV025B015, DV025D006, DV025D007, DV025D008, DV025D009, DV025D010, DV025E003, DV025E004, DV025E005, DV025E006, DV025E007, DV025E008, DV025E009, DV025E010, DV025E011, DV025E012, DV025E013, DV025E014, DV025E016, DV025E017, DV025F012, DV025F013, DV025F014, DV025F015, DV025F016, DV025G011, DV025G012, DV025G013, DV025G014, DV025G015, DV025G016, DV025G017, DV025G018, DV025G019, DV025G020, DV025J021, DV025J022, DV025J023, DV025J025, DV025J026, DV025J027, DV025J028, DV025J034, DV025J035, DV025J036, DV025J037, DV025J038, DV025J040, DV025K018, DV025K019, DV025K020, DV025K021, DV025K023, DV025K024, DV025K025, DV025K027, DV025M031, DV025M032, DV025M033, DV025M034, DV026B001, DV026B002, DV026B003, DV026B004, DV026B005, DV026B010, DV026B014, DV

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-23 05:30 UTC. Current status: ongoing.

Other recalls by Lutronic

All Lutronic recalls

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3110-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.