Atellica IM CA 19-9 (50 Test)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human seru…

On 2024-09-18 the FDA classified a Class II recall of Atellica IM CA 19-9 (50 Test)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human seru… by Siemens Healthcare Diagnostics, citing positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-3100-2024
ClassificationClass II
Statusongoing
Initiated2024-07-26
Published2024-09-18
Last updated2026-09-13 11:57 UTC
CompanySiemens Healthcare Diagnostics
DistributionUS

Product

Atellica IM CA 19-9 (50 Test)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human serum using the Atellica CI and Atellica IM Analyzers SMN: 10995490

Reason

Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample results above a concentration of 110 U/mL.

Identifiers

code_infoUDI-DI: (01)00630414598178(10)55974535(17)20240914 Kit Lot: 55974535, 55975535