5008X Standard HD Pre-Flush Standard Blood Tubing Set Part Number 03-5110-6

On 2026-09-09 the FDA classified a Class II recall of 5008X Standard HD Pre-Flush Standard Blood Tubing Set Part Number 03-5110-6 by Fresenius Medical Care Renal Therapies Group, citing heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovo….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-3084-2026
ClassificationClass II
Statusongoing
Initiated2026-08-10
Published2026-09-09
Last updated2026-09-15 05:30 UTC
CompanyFresenius Medical Care Renal Therapies Group
DistributionUS

Product

5008X Standard HD Pre-Flush Standard Blood Tubing Set Part Number 03-5110-6

Reason

Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.

Identifiers

code_infoLots 25KR01227 25NR01257 UDI-DI 00840861102556

Status history