FDA Device recall Z-3082-2024
Baxter EXACTAMIX Inlet, Non-Vented, High Volume Inlet, REF H938173, for use in a pharmaceutical compounding device
- FDA Device
- Class I
- ongoing
- Published 2024-09-25
On 2024-09-25 the FDA classified a Class I recall of Baxter EXACTAMIX Inlet, Non-Vented, High Volume Inlet, REF H938173, for use in a pharmaceutical compounding device by Baxter Healthcare, citing baxter Healthcare Corporation has received increased customer reports of particulate matter in the Automated Compounding Device Inlets (disposable inlet), Product Codes H938173, H….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-3082-2024 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2024-08-20 |
| Published | 2024-09-25 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Baxter Healthcare |
| Distribution | n/a |
Product
Baxter EXACTAMIX Inlet, Non-Vented, High Volume Inlet, REF H938173, for use in a pharmaceutical compounding device
Reason
Baxter Healthcare Corporation has received increased customer reports of particulate matter in the Automated Compounding Device Inlets (disposable inlet), Product Codes H938173, H938174, H938175, H938176, used with the ExactaMix and ExactaMix Pro compounders. Particulate matter has been observed within the inlet primary packaging inlet components, including within the sterile fluid path tubing, before use.
Identifiers
| code_info | UDI/DI 00085412475783, Lot Numbers: 803806 and lower |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3082-2024%22