FDA Device recall Z-3080-2017

AirLife(TM) Reduced Heparin ARTERIAL BLOOD SAMPLER, REF 9025RHTR, 1ML SYRINE 25G 5/8", LUER SLIP PROTECTIVE SHEATH, 3.2 IU, STERILE R, Rx O…

On 2017-09-06 the FDA classified a Class II recall of AirLife(TM) Reduced Heparin ARTERIAL BLOOD SAMPLER, REF 9025RHTR, 1ML SYRINE 25G 5/8", LUER SLIP PROTECTIVE SHEATH, 3.2 IU, STERILE R, Rx O… by Vyaire Medical, citing lot #0001047501 of the 9025RHTR ABG sampling kits have been manufactured containing the incorrect sub-assembly which includes a larger needle (23 Gage, 1" needle), larger syringe….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-3080-2017
ClassificationClass II
Statusterminated
Initiated2017-07-18
Published2017-09-06
Last updated2026-09-13 11:54 UTC
CompanyVyaire Medical
DistributionUS

Product

AirLife(TM) Reduced Heparin ARTERIAL BLOOD SAMPLER, REF 9025RHTR, 1ML SYRINE 25G 5/8", LUER SLIP PROTECTIVE SHEATH, 3.2 IU, STERILE R, Rx ONLY, CAREFUSION Lot #0001047501 of the 9025RHTR ABG sampling kits have been manufactured containing the incorrect sub-assembly which includes a larger needle (23 Gage, 1" needle), larger syringe (3mL), and larger amount of heparin (113IU), than what is intended for the kit.

Reason

Lot #0001047501 of the 9025RHTR ABG sampling kits have been manufactured containing the incorrect sub-assembly which includes a larger needle (23 Gage, 1" needle), larger syringe (3mL), and larger amount of heparin (113IU), than what is intended for the kit.

Identifiers

code_infoLot Number 0001047501

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3080-2017%22