FDA Device recall Z-3064-2026
HEALON EndoCoat PRO, REF: 10300012 (US/CAN) HEALON DUET PRO II, REF: 10300016 (US/CAN) HEALON EndoCoat PRO, REF: 10301012 (EU)
- FDA Device
- Class II
- ongoing
- Published 2026-09-09
On 2026-09-09 the FDA classified a Class II recall of HEALON EndoCoat PRO, REF: 10300012 (US/CAN) HEALON DUET PRO II, REF: 10300016 (US/CAN) HEALON EndoCoat PRO, REF: 10301012 (EU) by Johnson And Johnson Surgical Vision, citing surgical instructions sent for viscoelastic products used during ophthalmic viscosurgical cataract procedures including in part: If resistance is encountered during insertion, wit….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-3064-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2026-07-31 |
| Published | 2026-09-09 |
| Last updated | 2026-09-15 05:30 UTC |
| Company | Johnson And Johnson Surgical Vision |
| Distribution | US |
Product
HEALON EndoCoat PRO, REF: 10300012 (US/CAN) HEALON DUET PRO II, REF: 10300016 (US/CAN) HEALON EndoCoat PRO, REF: 10301012 (EU); HEALON DUET PRO II, REF: 10301016 (EU)
Reason
Surgical instructions sent for viscoelastic products used during ophthalmic viscosurgical cataract procedures including in part: If resistance is encountered during insertion, withdraw cannula and reassess the insertion conditions (i.e. insertion angle, incision size, etc.) before proceeding because forcing the cannula may cause corneal abrasion, iris trauma or descemet's membrane detachment.
Identifiers
| code_info | REF/UDI-DI/Lot: 10300012/5050474755093/UA32232, UA32250, UA32268, UA32290, UA32292, UA32303, UA32371, UP31203, UP31205;…REF/UDI-DI/Lot: 10300012/5050474755093/UA32232, UA32250, UA32268, UA32290, UA32292, UA32303, UA32371, UP31203, UP31205; 10300016/5050474755116/UA32447, UA32448, UA32449, UB33008, UB33026, UB33082, UB33083, UB33175; 10301012/5050474755109/UB33014, UB33041, UB33057, UB33060, UB33100, UB33190; 10301016/5050474755123/UB33084, UB33194 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-15 05:30 UTC. Current status: ongoing.
Other recalls by Johnson And Johnson Surgical Vision
- Johnson & Johnson VISION Reusable LAMINAR High Flow Irrigation Sleeve and Test Chamber 21 Gauge REF OPOHF21L The irrigation sleeve is a dev…
FDA Device · Class II · 2024-05-08 · Johnson And Johnson Surgical Vision - TECNIS Toric II OptiBlue IOL Models ZCW
FDA Device · Class II · 2024-01-24 · Johnson And Johnson Surgical Vision - TECNIS Symfony, Extended Range of Vision IOL, REF: ZXR00, STERILE.EO, Rx Only
FDA Device · Class II · 2023-08-09 · Johnson And Johnson Surgical Vision - VERITAS Advanced Infusion Packs (VRT-AI) and VERITAS Advanced Fluidics Packs (VRT-AF)
FDA Device · Class II · 2023-08-09 · Johnson And Johnson Surgical Vision - TECNIS Eyhance IOL, REF: ICB00, STERILE.EO, Rx Only
FDA Device · Class II · 2023-08-09 · Johnson And Johnson Surgical Vision
All Johnson And Johnson Surgical Vision recalls
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3064-2026%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.