On 2017-08-30 the FDA classified a Class II recall of ORA System with VerifEye+ Cart, Catalog Number 8065998307 For use during intraocular lens surgery by Alcon Research, citing some ORA Carts have the potential to return an incorrect IOL power measurement during cataract surgery. This issue appears to have been caused by a software coding error that resu….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-3049-2017 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2017-06-30 |
| Published | 2017-08-30 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Alcon Research |
| Distribution | US |
Product
ORA System with VerifEye+ Cart, Catalog Number 8065998307 For use during intraocular lens surgery
Reason
Some ORA Carts have the potential to return an incorrect IOL power measurement during cataract surgery. This issue appears to have been caused by a software coding error that results in the lens coefficients for an IOL model being downloaded from the Alcon server in an incorrect order.
Identifiers
| code_info | Serial numbers: C5013,C5014,C5015,C5016,C5017,C5018,C5019,C5020,C5021,C5023,C5024,C5025,C5026,C5027,C5028,C5029,C5030,…Serial numbers: C5013,C5014,C5015,C5016,C5017,C5018,C5019,C5020,C5021,C5023,C5024,C5025,C5026,C5027,C5028,C5029,C5030,C5031,C5032,C5033,C5034,C5035,C5036,C5038,C5039,C5040,C5041,C5042,C5043,C5044, C5045,C5046,C5047,C5048,C5049,C5050,C5051,C5053,C5054,C5055,C5059,C5060,C5061,C5062,C5064, C5065,C5066,C5069,C5071,C5075,C5076,C5077,C5078,C5079,C5080,C5081,C5082,C5083,C5090,C5093, C5094,C5097,C5098,C5101,C5102,C5105,C5106,C5107,C5108,C5109,C5110,C5111,C5112,C5113,C5116, C5117,C5119,C5120,C5121,C5122,C5125,C5126,C5127,C5128,C5129,C5130,C5131,C5132,C5133,C5134, C5136,C5137,C5138,C5139,C5140,C5141,C5142,C5143,C5144,C5146,C5147,C5148,C5149,C5150,C5151, C5152,C5153,C5154,C5155,C5156,C5157,C5158,C5159,C5160,C5161,C5163,C5164,C5165,C5166,C5167, C5168,C5169,C5170,C5171,C5173,C5174,C5175,C5176,C5177,C5178,C5179,C5180,C5181,C5182,C5183, C5184,C5185,C5186,C5187,C5188,C5189,C5190,C5191,C5192,C5193,C5194,C5195,C5196,C5197,C5198, C5199,C5200,C5201,C5202,C5203,C5204,C5205,C5207,C5214,C5215,C5216 |
|---|