FDA Device recall Z-3047-2017

VASER Amplifier, VASER 2.2 VASERlipo System, Rx Only, 115/230V - 60/50Hz 300VA The Vaser amplifier is intended for the fragmentation and em…

On 2017-08-30 the FDA classified a Class II recall of VASER Amplifier, VASER 2.2 VASERlipo System, Rx Only, 115/230V - 60/50Hz 300VA The Vaser amplifier is intended for the fragmentation and em… by Solta Medical, citing potential inaccurate calibration by service depot repair, which could contribute to patients being burnt by the device. This only impacts units serviced between the dates of 09/08….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-3047-2017
ClassificationClass II
Statusterminated
Initiated2017-06-01
Published2017-08-30
Last updated2026-09-13 11:54 UTC
CompanySolta Medical
DistributionAL, CA, CO, FL, NY, PA, TX

Product

VASER Amplifier, VASER 2.2 VASERlipo System, Rx Only, 115/230V - 60/50Hz 300VA The Vaser amplifier is intended for the fragmentation and emulsification of subcutaneous fatty tissues for aesthetic body contouring (modification of the anatomy).

Reason

Potential inaccurate calibration by service depot repair, which could contribute to patients being burnt by the device. This only impacts units serviced between the dates of 09/08/2016 and 05/23/2017.

Identifiers

code_infoReference No. 110-0021; Shipping P/N P007021-03; Device P/N P006982-03; Serial No. VAS10-0134, VAS13-0440, VAS12-0203, 000TWY.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.