FDA Device recall Z-3041-2018
Vascular Solutions Venture Rx Catheter, 6F, resterilized
- FDA Device
- Class II
- terminated
- Published 2018-09-19
On 2018-09-19 the FDA classified a Class II recall of Vascular Solutions Venture Rx Catheter, 6F, resterilized by Sps Sterilization, citing the expiration date on the device labeling exceeds its actual validated shelf life because the sterilization process used does not have adequate support for the extended shelf life.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-3041-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-03-07 |
| Published | 2018-09-19 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Sps Sterilization |
| Distribution | AZ, CA, FL, OH, PA, TX, VA, WV |
Product
Vascular Solutions Venture Rx Catheter, 6F, resterilized.
Reason
The expiration date on the device labeling exceeds its actual validated shelf life because the sterilization process used does not have adequate support for the extended shelf life.
Identifiers
| code_info | Code Number 5820-R; Lot Number 570183 (Serial Number 99804132016-07); Lot Number 583022 (Serial Number 17304202017-07, 17304202017-08) |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3041-2018%22