FDA Device recall Z-3040-2020

iCup Dx14 Drug Screen Cup(AMT Item Number/ AMT Description): X11-I-DXA-2147-011/ "iCup" Dx 14 Drug Screen Cup 14 Drug Detection: (THC50/COC…

On 2020-10-07 the FDA classified a Class II recall of iCup Dx14 Drug Screen Cup(AMT Item Number/ AMT Description): X11-I-DXA-2147-011/ "iCup" Dx 14 Drug Screen Cup 14 Drug Detection: (THC50/COC… by Ameditech, citing various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-3040-2020
ClassificationClass II
Statusterminated
Initiated2020-08-18
Published2020-10-07
Last updated2026-09-13 11:55 UTC
CompanyAmeditech
DistributionUS

Product

iCup Dx14 Drug Screen Cup(AMT Item Number/ AMT Description): X11-I-DXA-2147-011/ "iCup" Dx 14 Drug Screen Cup 14 Drug Detection: (THC50/COC150)+(MOP300/MET500)+(AMP500/BZO300)+ (BAR300/MTD300)+(BUPG10/TCA1000)+(MDMA500/OXY100)+(PCP25/PPX300) 5 Adulteration Tests: Cr, Ni, pH, Bl, S.G.", X11-I-DX-1147-022 /"iCup" Dx 14 Drug Screen Cup 14 Drug Detection: (THC50/COC150)+(MOP300/MET500)+(AMP500/BZO300)+ (BAR300/MTD300)+(BUPG10/TCA1000)+(MDMA500/OXY100)+(PCP25/PPX300)"

Reason

Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.

Identifiers

code_infoALL Lots Within Expiry

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.