FDA Device recall Z-3040-2017

HomeChoice Automated PD system Model numbers: 5C4471, 5C4471R Product Usage: The HomeChoice Automated Personal Cycler is intended for autom…

On 2017-08-23 the FDA classified a Class II recall of HomeChoice Automated PD system Model numbers: 5C4471, 5C4471R Product Usage: The HomeChoice Automated Personal Cycler is intended for autom… by Baxter Healthcare, citing baxter Healthcare Corporation has been made aware that users may not be following the instructions in the Operators Manual and incorrectly (using sharp objects) opening disposabl….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-3040-2017
ClassificationClass II
Statusterminated
Initiated2017-07-05
Published2017-08-23
Last updated2026-09-13 11:54 UTC
CompanyBaxter Healthcare
Distributionn/a

Product

HomeChoice Automated PD system Model numbers: 5C4471, 5C4471R Product Usage: The HomeChoice Automated Personal Cycler is intended for automatic control of dialysate solution exchanges in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.

Reason

Baxter Healthcare Corporation has been made aware that users may not be following the instructions in the Operators Manual and incorrectly (using sharp objects) opening disposable set packaging while setting up their Peritoneal Dialysis (PD) therapy, damaging the cassettes for the HomeChoice or HomeChoice PRO cyclers. If damaged cassettes are used, the cyclers may not consistently detect very small holes/cuts in the sheeting of the cassette in the patient valve region and the cycler may deliver air into the patient.