RecallRadar

FDA Device recall Z-3036-2020

ImmuTest Drugs of Abuse Cup (AMT Item Number/ AMT Description): IMCA-5AB-W / "ImmuTest Drugs of Abuse Cup AMP1000+BZO300+COC300+OPI2000+THC…

On 2020-10-07 the FDA classified a Class II recall of ImmuTest Drugs of Abuse Cup (AMT Item Number/ AMT Description): IMCA-5AB-W / "ImmuTest Drugs of Abuse Cup AMP1000+BZO300+COC300+OPI2000+THC… by Ameditech, citing various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-3036-2020
ClassificationClass II
Statusterminated
Initiated2020-08-18
Published2020-10-07
Last updated2026-09-13 11:55 UTC
CompanyAmeditech
DistributionUS

Product

ImmuTest Drugs of Abuse Cup (AMT Item Number/ AMT Description): IMCA-5AB-W / "ImmuTest Drugs of Abuse Cup AMP1000+BZO300+COC300+OPI2000+THC50 With Cr, Ni, pH, Bl, S.G.", IMC-8BBCE/ "ImmuTest Drugs of Abuse Cup 8 Drug Detection: (COC300/THC50)+AMP1000+BZO300+BUP10+MET500+MTD300+OPI300 NO ADULTERATION", IMCA-8CM5 /"ImmuTest Drugs of Abuse Cup (OXY100/MDMA500)+(PCP25/AMP1000)+COC150+MET500+OPI2000+THC50 with Cr, Ni, pH, Bl, S.G."

Reason

Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.

Identifiers

code_infoALL Lots Within Expiry

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3036-2020%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.