RecallRadar

FDA Device recall Z-3033-2026

ORT200, operating room table, including the ORT200 Covered Table Assembly, Reconditioned, Part Number 113821-600 and the ORT200 50/60Hz Cov…

On 2026-09-02 the FDA classified a Class II recall of ORT200, operating room table, including the ORT200 Covered Table Assembly, Reconditioned, Part Number 113821-600 and the ORT200 50/60Hz Cov… by Imris Imaging, citing the linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. Thi….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-3033-2026
ClassificationClass II
Statusongoing
Initiated2026-08-06
Published2026-09-02
Last updated2026-09-13 11:58 UTC
CompanyImris Imaging
DistributionUS

Product

ORT200, operating room table, including the ORT200 Covered Table Assembly, Reconditioned, Part Number 113821-600 and the ORT200 50/60Hz Covered Table Assy V2 Part Number 114148-000

Reason

The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure.

Identifiers

code_info113821-600 - Serial Numbers: 10003754, 10006997; 114148-000 - UDI-DI: 00857534006349, Serial Numbers: 10005189, 10005190, 10004463, 10004997

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.