RecallRadar

FDA Device recall Z-3032-2017

ThermaSure(TM), a CENORIN(TM) product, SERIES 130 MEDICAL DEVICE DRYER: (a) Model 135

On 2017-08-23 the FDA classified a Class II recall of ThermaSure(TM), a CENORIN(TM) product, SERIES 130 MEDICAL DEVICE DRYER: (a) Model 135 by Cenorin, citing dryer unit not heating. Manufacturing issue causes the terminal block to overheat, char and/or burn.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-3032-2017
ClassificationClass II
Statusterminated
Initiated2017-06-07
Published2017-08-23
Last updated2026-09-13 11:54 UTC
CompanyCenorin
DistributionUS

Product

ThermaSure(TM), a CENORIN(TM) product, SERIES 130 MEDICAL DEVICE DRYER: (a) Model 135; (b) Model 136 Forced air dryer for medical devices

Reason

Dryer unit not heating. Manufacturing issue causes the terminal block to overheat, char and/or burn.

Identifiers

code_info
(a) Serial Numbers: 350423, 350424, 351506, 351507, 351512, 351513, 351514, 351516, 351602, 351604, 351605, 351606, 35…(a) Serial Numbers: 350423, 350424, 351506, 351507, 351512, 351513, 351514, 351516, 351602, 351604, 351605, 351606, 351609, 351611, 351613, 351614, 351615, 351616, 351617, 351618, 351619, 351620, 351621, 351622, 351623, 351624 (b) Serial numbers: 361505, 361511, 361515, 361601, 361603, 361607, 361608, 361612

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3032-2017%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.