RecallRadar

FDA Device recall Z-3013-2018

CINtec p16 Histology (250) CE, Catalog Number 06695256001, model 825-4713 Immunohistochemistry (IHC) for in vitro diagnostic use

On 2018-09-26 the FDA classified a Class I recall of CINtec p16 Histology (250) CE, Catalog Number 06695256001, model 825-4713 Immunohistochemistry (IHC) for in vitro diagnostic use by Ventana Medical Systems, citing leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-3013-2018
ClassificationClass I
Statusterminated
Initiated2018-08-02
Published2018-09-26
Last updated2026-09-13 11:54 UTC
CompanyVentana Medical Systems
Distributionn/a

Product

CINtec p16 Histology (250) CE, Catalog Number 06695256001, model 825-4713 Immunohistochemistry (IHC) for in vitro diagnostic use.

Reason

Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.

Identifiers

code_infoLot Numbers: Y16507, Y23040, UDI: 4015630984398

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3013-2018%22