FDA Device recall Z-2999-2020
Speed Stitch Needle Cassette
- FDA Device
- Class II
- ongoing
- Published 2020-09-30
On 2020-09-30 the FDA classified a Class II recall of Speed Stitch Needle Cassette by Arthrocare, citing due to a supplier error where a portion of the raw material was inadvertently mixed. As a result, some needles may be made of commercially pure titanium as opposed to 304 stainles….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2999-2020 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2020-08-17 |
| Published | 2020-09-30 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Arthrocare |
| Distribution | AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX, UT, VA, WA, WI, WV |
Product
Speed Stitch Needle Cassette
Reason
Due to a supplier error where a portion of the raw material was inadvertently mixed. As a result, some needles may be made of commercially pure titanium as opposed to 304 stainless steel which does not meet the specification and could potentially break during use.
Identifiers
| code_info | Model: OM-8850 Lot Numbers: 2018434, 2018934, 2019558, 2019559, 2019929, 2019930, 2020931, 2022111, 2022641, 2030008,…Model: OM-8850 Lot Numbers: 2018434, 2018934, 2019558, 2019559, 2019929, 2019930, 2020931, 2022111, 2022641, 2030008, 2032212, 2033815, 2037604, 2036941, 2038701, 2039699, 2040500, 2042295, 2044957, 2047483, 2048575, 2051239 & 2032526 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2999-2020%22