Imactis Patient Fiducial (Patient Referential), REF: IHR-YY-XXX component used in conjunction with the ImactisCT-Navigation Workstation (RE…

On 2020-09-23 the FDA classified a Class II recall of Imactis Patient Fiducial (Patient Referential), REF: IHR-YY-XXX component used in conjunction with the ImactisCT-Navigation Workstation (RE… by Imactis, citing the firm has received reports concerning the patient fiducial disk separating from the body and users attempting to re-assemble the device. The use of a re-assembled patient fiduc….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2972-2020
ClassificationClass II
Statusongoing
Initiated2020-08-24
Published2020-09-23
Last updated2026-09-13 11:55 UTC
CompanyImactis
DistributionUS

Product

Imactis Patient Fiducial (Patient Referential), REF: IHR-YY-XXX component used in conjunction with the ImactisCT-Navigation Workstation (REF J02000/J00180)

Reason

The firm has received reports concerning the patient fiducial disk separating from the body and users attempting to re-assemble the device. The use of a re-assembled patient fiducial could lead to inability to navigate or inaccuracies such as improper needle trajectory.

Identifiers

code_infoAll Serial Numbers

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: ongoing.