FDA Device recall Z-2972-2018
DORO LUCENT Transitional Member, REF 1101.026 Component of headrest system used in open and percutaneous craniotomies and spinal surgeries
- FDA Device
- Class I
- terminated
- Published 2018-09-26
On 2018-09-26 the FDA classified a Class I recall of DORO LUCENT Transitional Member, REF 1101.026 Component of headrest system used in open and percutaneous craniotomies and spinal surgeries by Pro Med Instruments, citing the combination of two different design revisions of DORO LUCENT Base Unit, Locking Transitional Member, and Transitional Member within one set up might lead to a malfunction and….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2972-2018 |
| Classification | Class I |
| Status | terminated |
| Initiated | 2018-07-05 |
| Published | 2018-09-26 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Pro Med Instruments |
| Distribution | CA, GA, MS, OH, TN, TX, WV |
Product
DORO LUCENT Transitional Member, REF 1101.026 Component of headrest system used in open and percutaneous craniotomies and spinal surgeries
Reason
The combination of two different design revisions of DORO LUCENT Base Unit, Locking Transitional Member, and Transitional Member within one set up might lead to a malfunction and failure of the headrest system.
Identifiers
| code_info | Serial Numbers: 1704002 1704004 1704005 1704007 1704011 1704041 1704010 1704019 1704012 1704014 1704016 170…Serial Numbers: 1704002 1704004 1704005 1704007 1704011 1704041 1704010 1704019 1704012 1704014 1704016 1704017 1704021 1704024 1704025 1704036 1704038 1704018 1704008 1704031 1704026 1704027 1704028 1704029 1704030 1704032 1704034 1704035 1704006 1704037 1704039 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2972-2018%22