FDA Device recall Z-2970-2020
Siemens epoc BGEM Test Card-In vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial…
- FDA Device
- Class II
- terminated
- Published 2020-09-23
On 2020-09-23 the FDA classified a Class II recall of Siemens epoc BGEM Test Card-In vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial… by Siemens Healthcare Diagnostics, citing sporadically inconsistent discrepant (low bias) glucose results on card lot 01-20095-10. Potential exists for a delay in diagnosis of hyperglycemia or unnecessary treatment for hy….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2970-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2020-08-28 |
| Published | 2020-09-23 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Siemens Healthcare Diagnostics |
| Distribution | US |
Product
Siemens epoc BGEM Test Card-In vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care SMN#/Model #: 10736382
Reason
Sporadically inconsistent discrepant (low bias) glucose results on card lot 01-20095-10. Potential exists for a delay in diagnosis of hyperglycemia or unnecessary treatment for hypoglycemia.
Identifiers
| code_info | Lot Number: 01-20095-10 |
|---|