RecallRadar

FDA Device recall Z-2962-2024

Impella 5.5 with SmartAssist Set, CA

On 2024-09-11 the FDA classified a Class II recall of Impella 5.5 with SmartAssist Set, CA by Abiomed, citing A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2962-2024
ClassificationClass II
Statusongoing
Initiated2024-08-05
Published2024-09-11
Last updated2026-09-13 11:57 UTC
CompanyAbiomed
DistributionUS

Product

Impella 5.5 with SmartAssist Set, CA; Product Code: 0550-0004;

Reason

A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.

Identifiers

code_info
Product Code: 0550-0004; UDI-DI: 00813502010466; Serial Numbers: 439898 437211A 437204A 440938 439938 439939 453784 453…Product Code: 0550-0004; UDI-DI: 00813502010466; Serial Numbers: 439898 437211A 437204A 440938 439938 439939 453784 453785 476355 476356 470678 493496 493497 491749 491752 491751 478569A 483483A 491750 485277A 485281A; Batch Numbers: 2024337828 2024338391 2024338394 2024340273 2024341579 2024341586 2024356969 2024356971 2024397878 2024397886 2024399646 2025440057 2025440061 2025448005 2025452639 2025452668 2025484262 2025484276 2025448006 2025500661 2025500668;

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2962-2024%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.